DOI 10.1097/jtn.0000000000000010
Full of sound and fury, signifying nothing: burden of transient noncritical monitor alarms in a trauma resuscitation unit.
Аннотация
We examined the types of patient monitor alarms encountered in the trauma resuscitation unit of a major level 1 trauma center. Over a 1-year period, 316688 alarms were recorded for 6701 trauma patients (47 alarms/patient). Alarms were more frequent among patients with a Glasgow Coma Scale of 8 or less. Only 2.4% of all alarms were classified as "patient crisis," with the rest in the presumably less critical categories "patient advisory," "patient warning," and "system warning." Nearly half of alarms were ≤5 seconds in duration. In this patient population, a 2-second delay would reduce alarms by 25%, and a delay of 5 seconds would reduce all alarms by 49%.