MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1159/000501256

Медицинская библиотека · Medical principles and practice : international journal of the Kuwait University, Health Science Centre · 2019

Multimodal Intraoperative Spinal Cord Monitoring during Spinal Deformity Surgery: Efficacy, Diagnostic Characteristics, and Algorithm Development.

Tsirikos Athanasios I, Duckworth Andrew D, Henderson Lindsay E, Michaelson Ciara

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

This study aims to present the diagnostic characteristics of multimodal intraoperative monitoring (IOM) in spinal deformity surgery and to define and categorise the neuromonitoring events, as well as propose an algorithm of action.

MATERIALS AND

METHODS

We reviewed 1,155 consecutive patients (807 female, 348 male) who underwent deformity correction using standardised perioperative care, cortical/cervical somatosensory evoked potentials (SSEPs), and upper/lower limb transcranial electrical motor evoked potential (MEPs) by a single surgeon. The mean age at surgery was 13.8 years (range 10-23.3). We categorised IOM events as true, transient true, and false positive or negative. Diagnostic performance criteria were calculated.

RESULTS

The most common diagnosis was adolescent idiopathic scoliosis in 717 (62%) patients. We identified 3 true positive monitoring events occurring in 2 patients (0.17%), 8 transient true positive (0.69%), and 8 transient false positive events (0.69%). There were no false negative events and no patient had postoperative neurological complications. The multimodal IOM technique had a sensitivity of 100%, specificity of 99.3%, positive predictive value of 55.6%, and negative predictive value of 100%. Sensitivity was 100% for MEPs and multimodal monitoring compared to 20% for cortical or cervical SSEPs. The frequency of true or transient true positive events was higher (p = 0.07) in Scheuermann's kyphosis (3/91 patients, 3.3%) compared to adolescent idiopathic scoliosis (6/717 patients, 0.84%).

CONCLUSION

Multimodal IOM is highly sensitive and specific for spinal cord injury. This technique is reliable for the assessment of the condition of the spinal cord during major deformity surgery. We propose an algorithm of intraoperative action to allow close cooperation between the surgical, anaesthetic, and neurophysiology teams and to prevent neurological deficits.

Для цитирования: Tsirikos Athanasios I, Duckworth Andrew D, Henderson Lindsay E, Michaelson Ciara. Multimodal Intraoperative Spinal Cord Monitoring during Spinal Deformity Surgery: Efficacy, Diagnostic Characteristics, and Algorithm Development.. Medical principles and practice : international journal of the Kuwait University, Health Science Centre. 2019. DOI: 10.1159/000501256.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.