Improving atrial fibrillation or flutter detection and management by smartphone-based photoplethysmography rhythm monitoring following cardiac surgery: a pragmatic randomized trial.
Аннотация
AIMS: Atrial fibrillation (AF) and atrial flutter (AFL) after cardiac surgery are common and associated with adverse outcomes. The increased risk related to AF or AFL may extend beyond discharge. This study aims to determine whether photoplethysmography (PPG)-based smartphone monitoring to detect AF or AFL after hospital discharge following cardiac surgery improves AF management.
METHODS AND
RESULTS
The intervention group performed 1 min rhythm checks three times daily using a smartphone-based PPG application during 6 weeks after hospitalization for cardiac surgery. The primary outcome involved AF management interventions by independent physicians, including initiation of oral anticoagulation (OAC), direct cardioversion, and up-titration or initiation of antiarrhythmic drugs. The study included 450 patients [mean (SD) age, 64.1 (9.2) years; 96 women (21.3%); 130 patients with AF history (28.9%); median (IQR) CHA2DS2-VASc score, 2 (1-3)], of whom 238 were randomized to PPG-based monitoring and 212 to usual care. AF/AFL was detected with PPG or electrocardiography in 44 patients (18.5%) in the monitoring group and 4 patients (1.9%) in the usual care group (OR 11.8; 95% CI, 4.2-33.3; P < 0.001); these were new detections in, respectively, 22 patients (9.2%) and 1 patient (0.5%) (OR 21.3; 95% CI, 2.9-166.7; P = 0.003). AF management interventions occurred in 24 patients (10.1%) in the monitoring group compared to 5 patients (2.4%) in the usual care group [odds ratio (OR), 5.1; 95% CI, 1.8-14.4; P = 0.002].
CONCLUSION
In unselected patients discharged home following cardiac surgery, PPG-based smartphone monitoring revealed significantly more AF/AFL which led to significantly more optimization of AF management.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.