Experimental comparison of mesenteric vessel sealing and thermal damage between one bipolar and two ultrasonic shears devices.

Noble E J, Smart N J, Challand C, Sleigh K, Oriolowo A, Hosie K B

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Several instruments are available for open and laparoscopic dissection, including electrothermal bipolar vessel sealers and ultrasonically coagulating shears. The vessel sealing ability of three devices in colorectal specimens was compared in an experimental study.

METHODS

Surgical specimens from patients scheduled for elective open or laparoscopic colorectal resection were allocated to one of the three devices. After removal of the surgical specimen, up to eight mesenteric vessels were dissected ex vivo and sealed using the allocated instrument. The vessel seal was tested for the maximum pressure at which it leaked and then assessed by a pathologist for depth of thermal tissue damage.

RESULTS

A total of 93 vessels from 18 patients were assessed ex vivo (LOTUS™ n = 33; Harmonic Ace® n = 30; LigaSure™ n = 30), a median of 6 (range 1-8) vessels per surgical specimen with a mean(s.d.) diameter of 1·06(0·70) mm and wall thickness of 0·29(0·19) mm. Mean(s.d.) bursting pressures were 1170(440), 1470(670) and 1510(740) mmHg with LOTUS(™) , Harmonic Ace® and LigaSure™ respectively. ANCOVA showed no difference in bursting pressure between the instruments (P = 0·058). The depth of thermal damage was significantly greater with LigaSure™ (3·37(1·44) mm) than with LOTUS(™) (2·18(0·99) mm; P < 0·001) or Harmonic Ace® (1·95(0·92) mm; P < 0·001).

CONCLUSION

All three instruments were equally good at sealing blood vessels, with bursting pressures well above physiological blood pressure levels.

REGISTRATION NUMBER: NCT01121614 (http://www.clinicaltrials.gov).

Для цитирования: Noble E J, Smart N J, Challand C, Sleigh K, Oriolowo A, Hosie K B. Experimental comparison of mesenteric vessel sealing and thermal damage between one bipolar and two ultrasonic shears devices.. The British journal of surgery. 2011. DOI: 10.1002/bjs.7433.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.