Nuclear myocardial perfusion imaging with a cadmium-telluride semiconductor detector gamma camera in patients with acute myocardial infarction.
Аннотация
OBJECTIVE
Since myocardial perfusion imaging (MPI) with conventional sodium iodine (NaI) device has low spatial resolution, there have been some cases in which small structures such as non-transmural myocardial infarction could not be properly detected. The purpose of this study was to evaluate potential usefulness of cadmium-telluride (CdTe) semiconductor detector-based high spatial resolution gamma cameras in detecting myocardial infarction sites, especially non-transmural infarction.
METHODS
A total of 38 patients (mean age ± SD: 64 ± 21 year) who were clinically diagnosed with acute myocardial infarction were included. Twenty-eight cases of them were with ST segment elevation myocardial infarction (STEMI) and 10 cases with non-ST segment elevation myocardial infarction (NSTEMI). In all patients, myocardial perfusion single photon emission computed tomography images were acquired with Infinia (NaI device) and R1-M (CdTe device), and the images were compared concerning the detectability of acute myocardial infarction sites.
RESULTS
The detection rates of the myocardial infarction site in cases with STEMI were 100% both by NaI and CdTe images. In cases with NSTEMI, detection rate by NaI images was 50%, while that of CdTe images was 100% (p = 0.033). The summed rest score (SRS) value derived from CdTe images was significantly higher than that from NaI images in cases with STEMI [NaI images: 12 (7-18) versus CdTe images: 14 (9-20)] (p < 0.001). SRS derived from CdTe images was significantly higher than that derived from NaI images in cases with NSTEMI [NaI images: 2 (0-5) versus CdTe images: 6 (6-8)] (p = 0.006).
CONCLUSIONS
These results indicate that MPI using CdTe-semiconductor device will provide a much more accurate assessment of acute myocardial infarction in comparison to current methods.