BACKGROUND

Custom-made fenestrated and branched endografts are considered the gold standard devices for thoracoabdominal repair, with lower perioperative mortality and morbidities than open surgical repair. Although they are not feasible in urgent and emergent settings, in which sandwich techniques are often necessary, custom-made devices can still be used as bail-out options to correct late complications such as high-flow endoleaks, as shown in this case report.

METHODS

A complex, symptomatic thoracoabdominal aneurysm, unsuitable for open repair, was treated by sandwich technique on the celiac trunk despite having a short distal neck. After the finding of a high-flow type IB endoleak during postoperative follow-up, reintervention was planned with a custom-made endograft with a distal scallop, to preserve patency of the superior mesenteric artery.

RESULTS

The procedure achieved technical success, with complete exclusion of the aneurysm and patency of all the visceral arteries.

CONCLUSIONS

Bail-out endovascular repair with custom-made endografts after failure of parallel graft repair of thoracoabdominal aneurysms is a valuable treatment option preventing the risks related to an open conversion.

Для цитирования: Marone Enrico Maria, Rinaldi Luigi Federico, Diaco Domenico Antonio, Argenteri Angelo. Failure of Sandwich Technique for Thoracoabdominal Aneurysm Treated with Custom-Made Fenestrated Endograft.. Annals of vascular surgery. 2018. DOI: 10.1016/j.avsg.2018.06.022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.