Diagnostic Accuracy of Spiral Whole-Heart Quantitative Adenosine Stress Cardiovascular Magnetic Resonance With Motion Compensated L1-SPIRIT.

Pan Jonathan A, Robinson Austin A, Yang Yang, Lozano Patricia Rodriguez, McHugh Stephen, Holland Eric M, Meyer Craig H, Taylor Angela M, Kramer Christopher M, Salerno Michael

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Variable density spiral (VDS) pulse sequences with motion compensated compressed sensing (MCCS) reconstruction allow for whole-heart quantitative assessment of myocardial perfusion but are not clinically validated.

PURPOSE: Assess performance of whole-heart VDS quantitative stress perfusion with MCCS to detect obstructive coronary artery disease (CAD).

STUDY TYPE: Prospective cross sectional.

POPULATION: Twenty-five patients with chest pain and known or suspected CAD and nine normal subjects.

FIELD STRENGTH/SEQUENCE: Segmented steady-state free precession (SSFP) sequence, segmented phase sensitive inversion recovery sequence for late gadolinium enhancement (LGE) imaging and VDS sequence at 1.5 T for rest and stress quantitative perfusion at eight short-axis locations.

ASSESSMENT: Stenosis was defined as ≥50% by quantitative coronary angiography (QCA). Visual and quantitative analysis of MRI data was compared to QCA. Quantitative analysis assessed average myocardial perfusion reserve (MPR), average stress myocardial blood flow (MBF), and lowest stress MBF of two contiguous myocardial segments. Ischemic burden was measured visually and quantitatively.

STATISTICAL TESTS: Student's t-test, McNemar's test, chi-square statistic, linear mixed-effects model, and area under receiver-operating characteristic curve (ROC).

RESULTS

Per-patient visual analysis demonstrated a sensitivity of 84% (95% confidence interval [CI], 60%-97%) and specificity of 83% [95% CI, 36%-100%]. There was no significant difference between per-vessel visual and quantitative analysis for sensitivity (69% [95% CI, 51%-84%] vs. 77% [95% CI, 60%-90%], P = 0.39) and specificity (88% [95% CI, 73%-96%] vs. 80% [95% CI, 64%-91%], P = 0.75). Per-vessel quantitative analysis ROC showed no significant difference (P = 0.06) between average MPR (0.68 [95% CI, 0.56-0.81]), average stress MBF (0.74 [95% CI, 0.63-0.86]), and lowest stress MBF (0.79 [95% CI, 0.69-0.90]). Visual and quantitative ischemic burden measurements were comparable (P = 0.85).

DATA

CONCLUSION

Whole-heart VDS stress perfusion demonstrated good diagnostic accuracy and ischemic burden evaluation. No significant difference was seen between visual and quantitative diagnostic performance and ischemic burden measurements.

EVIDENCE LEVEL: 2 TECHNICAL EFFICACY: Stage 2.

Для цитирования: Pan Jonathan A, Robinson Austin A, Yang Yang, Lozano Patricia Rodriguez, McHugh Stephen, Holland Eric M, Meyer Craig H, Taylor Angela M, Kramer Christopher M, Salerno Michael. Diagnostic Accuracy of Spiral Whole-Heart Quantitative Adenosine Stress Cardiovascular Magnetic Resonance With Motion Compensated L1-SPIRIT.. Journal of magnetic resonance imaging : JMRI. 2021. DOI: 10.1002/jmri.27620.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.