STUDY

OBJECTIVES

To determine the optimal Actiwatch 2 setting configuration for the estimation of total sleep time (TST) in persons with suspected idiopathic hypersomnia.

METHODS

Thirty-three patients with a diagnosis of idiopathic hypersomnia (28 female; mean age = 33.7 ± 10.5) underwent ad libitum polysomnography with concurrent use of the Actiwatch 2. Actiwatch 2 sleep-wake activity threshold (SWAT; Low, Medium, and High) and sleep immobility onset and offset (SIOO; 5, 10, 15, 20, 25, and 30 epoch) duration were modified during data processing. The resultant 18 unique setting combinations were subsequently evaluated using Bland-Altman and epoch comparison analyses to determine optimal settings relative to polysomnography.

RESULTS

Low SWAT + 25 Epoch SIOO displayed the least divergence from polysomnography (mean difference 3.4 minutes). Higher SWAT and lower SIOO increased sensitivity and accuracy, but at the expense of reducing specificity and the ability to accurately estimate TST.

CONCLUSIONS

These results demonstrate that actigraphic settings should be carefully considered when estimating sleep duration. The Low + 25 Epoch configuration is indicated as most optimal for estimating TST in persons with suspected idiopathic hypersomnia.

COMMENTARY: A commentary on this article appears in this issue on page 539.

Для цитирования: Cook Jesse D, Eftekari Sahand C, Leavitt Lydia A, Prairie Michael L, Plante David T. Optimizing Actigraphic Estimation of Sleep Duration in Suspected Idiopathic Hypersomnia.. Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine. 2019. DOI: 10.5664/jcsm.7722.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.