DOI 10.1177/02841851221100079

Медицинская библиотека · Acta radiologica (Stockholm, Sweden : 1987) · 2022

Spatial analysis of acoustic noise transfer function with a human-body phantom in a clinical MRI scanner.

Hamaguchi Takashi, Miyati Tosiaki, Ohno Naoki, Matsushita Tatsuhiko, Takata Tadanori, Matsuura Yukihiro, Kobayashi Satoshi, Gabata Toshifumi

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The acoustic noise in magnetic resonance imaging (MRI) potentially depends on the measurement position and presence of a patient inside the scanner bore.

PURPOSE: To analyze the spatial characteristics of the acoustic noise by using the gradient-pulse-to-acoustic-noise transfer function (GPAN-TF) with and without a human-body phantom on the examination table.

MATERIAL AND

METHODS

Acoustic noise waveforms were acquired at 80 and 110 measurement positions with and without a phantom. The GPAN-TFs µPa/(mT/m) in the coils were calculated by deconvolution. The phantom effect on the spatial distribution of the acoustic noise was assessed using the peak sound pressure levels (SPLs), mean values, peak values, and peak frequencies of the GPAN-TFs.

RESULTS

The peak SPLs in all positions for the X-, Y-, and Z-gradient coils were increased by 11.1 dB, 1.4 dB, and 6.1 dB, respectively, compared with the peak SPL of the magnetic isocenter. The maximum peak SPLs among all positions of the X-, Y-, and Z-gradient coils with the phantom were increased by 4.9 dB, 7.4 dB, and 6.9 dB, respectively, relative to those without the phantom. However, the peak SPLs decreased at some positions with the phantom placed on the table (X-gradient coil = 4.6 dB, Y-gradient coil = 5.0 dB, Z-gradient coil = 8.4 dB). The most common peak frequencies were in the range of 2000-3000 Hz.

CONCLUSION

"Hotspot" areas with and without the phantom were associated with acoustic noise sources in the clinical MRI scanner and were enhanced by the phantom's presence.

Для цитирования: Hamaguchi Takashi, Miyati Tosiaki, Ohno Naoki, Matsushita Tatsuhiko, Takata Tadanori, Matsuura Yukihiro, Kobayashi Satoshi, Gabata Toshifumi. Spatial analysis of acoustic noise transfer function with a human-body phantom in a clinical MRI scanner.. Acta radiologica (Stockholm, Sweden : 1987). 2022. DOI: 10.1177/02841851221100079.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.