Artifacts in fluoroscopy and changes in radiation dose caused by heating blankets and insulating covers during simulated endovascular treatment.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Podsiadło Paweł, Liszka Grzegorz, Popiela Tadeusz, Sanak Tomasz, Kosiński Sylweriusz, Darocha Tomasz
Аннотация
PURPOSE: We aimed to assess whether insulating covers and warming systems cause artifacts in fluoroscopy, and whether they alter the radiation dose.
METHODS: Eight insulating and warming systems were wrapped around the phantom in order to obtain images in fluoroscopy, and to measure the absorbed and scattered radiation dose. A dosimeter, endovascular catheters, and stents were placed into a phantom. The other dosimeter was placed outside of a C-arm table, at the operator's and anesthesiologist's locations.
RESULTS: Most of the insulating covers did not cause artifacts in the fluoroscopy and led to a significant decrease in both the absorbed and scattered radiation dose. The highest decrease in the absorbed dose was observed with metalized foil (- 2.09%; p = 0.001) and in the scattered dose with Helios cover (- 55%; p < 0.001). Only one heating system (Ready Heat combined with Hypothermia Prevention and Management Kit cover) caused significant artifacts and increased radiation up to 99% (p < 0.001).
CONCLUSION: Thermal insulation may be maintained during X-ray-guided emergency endovascular procedures in trauma victims. Self-heating blankets should be replaced with another warming system.
METHODS: Eight insulating and warming systems were wrapped around the phantom in order to obtain images in fluoroscopy, and to measure the absorbed and scattered radiation dose. A dosimeter, endovascular catheters, and stents were placed into a phantom. The other dosimeter was placed outside of a C-arm table, at the operator's and anesthesiologist's locations.
RESULTS: Most of the insulating covers did not cause artifacts in the fluoroscopy and led to a significant decrease in both the absorbed and scattered radiation dose. The highest decrease in the absorbed dose was observed with metalized foil (- 2.09%; p = 0.001) and in the scattered dose with Helios cover (- 55%; p < 0.001). Only one heating system (Ready Heat combined with Hypothermia Prevention and Management Kit cover) caused significant artifacts and increased radiation up to 99% (p < 0.001).
CONCLUSION: Thermal insulation may be maintained during X-ray-guided emergency endovascular procedures in trauma victims. Self-heating blankets should be replaced with another warming system.