PURPOSE: To evaluate the magnetic field interactions with Yasargil aneurysm clips exposed to a 1.5-T MR system.

MATERIALS AND

METHODS

Sixteen Yasargil aneurysm clips made with Phynox (10 clips returned to the manufacturer due to suspected magnetic field interactions and six clips obtained from the manufacturer in sterile packages) were tested for magnetic field interaction at 1.5 T by using techniques to assess magnetic field-induced translational (deflection angle test) and torque (changes in alignment or rotation) forces.

RESULTS

None of the aneurysm clips demonstrated magnetic field interactions; that is, deflection angles were zero and there were no changes in alignment or evidence of rotational forces present.

CONCLUSION

It is safe to perform MR procedures at 1.5 T or less in patients with Yasargil aneurysm clips made with nonferromagnetic material.

Для цитирования: Shellock F G, Kanal E. Yasargil aneurysm clips: evaluation of interactions with a 1.5-T MR system.. Radiology. 1998. DOI: 10.1148/radiology.207.3.9609877.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.