DOI 10.1097/sih.0b013e3182a833e0
A simulator-based study of in-flight auscultation.
Аннотация
INTRODUCTION
The use of a stethoscope is essential to the delivery of continuous, supportive en route care during aeromedical evacuations. We compared the capability of 2 stethoscopes (electronic, Litmann 3000; conventional, Litmann Cardiology III) at detecting pathologic heart and lung sounds, aboard a C135, a medical transport aircraft.
METHODS
Sounds were mimicked using a mannequin-based simulator SimMan. Five practitioners examined the mannequin during a fly, with a variety of abnormalities as follows: crackles, wheezing, right and left lung silence, as well as systolic, diastolic, and Austin-Flint murmur. The comparison for diagnosis assessed (correct or wrong) between using the electronic and conventional stethoscopes were performed as a McNemar test.
RESULTS
A total of 70 evaluations were performed. For cardiac sounds, diagnosis was right in 0/15 and 4/15 auscultations, respectively, with conventional and electronic stethoscopes (McNemar test, P = 0.13). For lung sounds, right diagnosis was found with conventional stethoscope in 10/20 auscultations versus 18/20 with electronic stethoscope (P = 0.013).
CONCLUSIONS
Flight practitioners involved in aeromedical evacuation on C135 plane are more able to practice lung auscultation on a mannequin with this amplified stethoscope than with the traditional one. No benefit was found for heart sounds.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.