DOI 10.1038/s41598-025-02349-5

Медицинская библиотека · Scientific reports · 2025

A noncontact vital sign sensor demonstrating a strong correlation with an electrocardiogram electrode and a CO2 sensor.

Nagano Tatsuya, Yamamoto Nobuyuki, Shinomiya Sae, Kato Hiroshi, Okamoto Mariko, Hazama Daisuke, Bo Ryosuke, Yamamura Tomohiko, Imai Takumi, Fukuzawa Koji, Funayama Takuya, Kano Koji, Katsurada Naoko

СтатьяПеревод в очереди

Аннотация

Accurate assessment of vital signs is important for reducing mortality. The aim of this study was to validate the effectiveness and safety of noncontact vital sign sensors. Interference tests were conducted with a noncontact vital sign sensor and medical devices. Inpatients' heart and respiratory rates were monitored via this sensor, and the measurements of this sensor were compared with those of reference medical equipment. Noncontact vital sign sensors and medical devices did not interfere with each other. A total of 21 patients (10 adults and 11 children, including 1 baby) were analysed. For all patients, the correlation coefficients for the HR and RR were 0.86 and 0.96, respectively. In adult patients, the correlation coefficients for the HR and RR were 0.75 and 0.96, respectively. In paediatric patients, the correlation coefficients for the HR and RR were 0.82 and 0.94, respectively. No effects of noncontact vital sign sensors on patients, surrounding patients or medical equipment were observed. Noncontact vital sign sensors are accurate and safe.

Для цитирования: Nagano Tatsuya, Yamamoto Nobuyuki, Shinomiya Sae, Kato Hiroshi, Okamoto Mariko, Hazama Daisuke, Bo Ryosuke, Yamamura Tomohiko, Imai Takumi, Fukuzawa Koji, Funayama Takuya, Kano Koji, Katsurada Naoko. A noncontact vital sign sensor demonstrating a strong correlation with an electrocardiogram electrode and a CO2 sensor.. Scientific reports. 2025. DOI: 10.1038/s41598-025-02349-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.