MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/nmo.14692

Медицинская библиотека · Neurogastroenterology and motility · 2023

Assessing mechanical function of peristalsis with functional lumen imaging probe panometry: Contraction power and displaced volume.

Halder Sourav, Pandolfino John E, Kahrilas Peter J, Koop Andree, Schauer Jacob, Araujo Isis K, Elisha Guy, Kou Wenjun, Patankar Neelesh A, Carlson Dustin A

СтатьяПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

BACKGROUND AND AIMS: The distal contractile integral (DCI) quantifies the contractile vigor of primary peristalsis on high-resolution manometry (HRM), whereas no such metric exists for secondary peristalsis on functional lumen imaging probe (FLIP) panometry. This study aimed to evaluate novel FLIP metrics of contraction power and displaced volume in asymptomatic controls and a patient cohort.

METHODS

Thirty-five asymptomatic controls and adult patients (with normal esophagogastric junction outflow/opening and without spasm) who completed HRM and FLIP panometry were included. The patient group also completed timed barium esophagram (TBE). Contraction power (estimate of esophageal work over time) and displaced volume (estimate of contraction-associated fluid flow) were computed from FLIP. HRM was analyzed per Chicago Classification v4.0.

KEY

RESULTS

In controls, median (5th-95th percentile) contraction power was 27 mW (10-44) and displaced volume was 43 mL (17-66). 95 patients were included: 72% with normal motility on HRM, 17% with ineffective esophageal motility (IEM), and 12% with absent contractility. Among patients, DCI was significantly correlated with both contraction power (rho = 0.499) and displaced volume (rho = 0.342); p values < 0.001. Both contraction power and displaced volume were greater in patients with normal motility versus IEM or absent contractility, complete versus incomplete bolus transit, and normal versus abnormal retention on TBE; p values < 0.02.

CONCLUSIONS

FLIP panometry metrics of contraction power and displaced volume appeared to effectively quantify peristaltic vigor. These novel metrics may enhance evaluation of esophageal motility with FLIP panometry and provide a reliable surrogate to DCI on HRM.

Для цитирования: Halder Sourav, Pandolfino John E, Kahrilas Peter J, Koop Andree, Schauer Jacob, Araujo Isis K, Elisha Guy, Kou Wenjun, Patankar Neelesh A, Carlson Dustin A. Assessing mechanical function of peristalsis with functional lumen imaging probe panometry: Contraction power and displaced volume.. Neurogastroenterology and motility. 2023. DOI: 10.1111/nmo.14692.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.