PURPOSE: To characterize the diffusion time-dependence in muscle in healthy adult volunteers, boys with Duchenne's muscular dystrophy (DMD), and age-matched controls in a clinically feasible acquisition time for pediatric applications.

METHODS

Diffusion data were acquired using a pulsed gradient stimulated echo diffusion preparation at 5 different diffusion times (70, 130, 190, 250, and 330 ms), at 4 different b-values (0, 200, 400, 600, and 800 s/mm2 ) and 6 directions (orthogonal x, y, and z and diagonal xy, xz, and yz) and processed to obtain standard diffusion indices (mean diffusivity [MD] and fractional anisotropy [FA]) at each diffusion time.

RESULTS

Time-dependent diffusion was seen in muscle in healthy adult volunteers, boys with DMD, and age-matched controls. Boys with DMD showed reduced MD and increased FA values in comparison to age matched controls across a range of diffusion times. A diffusion time of Δ = 190 ms had the largest effect size.

CONCLUSIONS

These results could be used to optimize diffusion imaging in this disease further and imply that these diffusion indices may become an important biomarker in monitoring progression in DMD in the future.

Для цитирования: McDowell Amy R, Feiweier Thorsten, Muntoni Francesco, Hall Matt G, Clark Chris A. Clinically feasible diffusion MRI in muscle: Time dependence and initial findings in Duchenne muscular dystrophy.. Magnetic resonance in medicine. 2021. DOI: 10.1002/mrm.28945.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.