Accuracy of ECG-based screening for sleep-disordered breathing: a survey of all male workers in a transport company.
Аннотация
PURPOSE: Sleep-disordered breathing (SDB) is associated with increased risk for cardiovascular morbidity and mortality and for sleepiness-related accidents, but >75 % of the patients remain undiagnosed. We sought to determine the diagnostic accuracy of ECG-based detection of SDB when used for population-based screening.
METHODS
All male workers, mostly truck drivers, of a transport company (n = 165; age, 43 ± 12 years) underwent standard attended overnight polysomnography. Cyclic variation of heart rate (CVHR), a characteristic pattern of heart rate associated with SDB, was detected from single-lead ECG signals during the polysomnography by a newly developed automated algorithm of autocorrelated wave detection with adaptive threshold (ACAT).
RESULTS
Among 165 subjects, the apnea-hypopnea index (AHI) was ≥5 in 62 (38 %), ≥15 in 26 (16 %), and ≥30 in 16 (10 %). The number of CVHR per hour (CVHR index) closely correlated with AHI [r = 0.868 (95 % CI, 0.825-0.901)]. The areas under the receiver operating characteristic curves for detecting subjects with AHI ≥5, ≥15, and ≥30 were 0.796 (95 % CI, 0.727-0.855), 0.974 (0.937-0.993), and 0.997 (0.971-0.999), respectively. With a predetermined criterion of CVHR index ≥15, subjects with AHI ≥15 were identified with 88 % sensitivity and 97 % specificity (likelihood ratios for positive and negative test, 30.7 and 0.12). The classification performance was retained in subgroups of subjects with obesity, hypertension, diabetes mellitus, dyslipidemia, and decreased autonomic function.
CONCLUSIONS
The CVHR obtained by the ACAT algorithm may provide a useful marker for screening for moderate-to-severe SDB among apparently healthy male workers.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.