m+Medelevator.RU
← Поиск публикаций
Journal of the American Heart Association · 2020

Intracoronary Saline-Induced Hyperemia During Coronary Thermodilution Measurements of Absolute Coronary Blood Flow: An Animal Mechanistic Study.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Adjedj Julien, Picard Fabien, Collet Carlos, Bruneval Patrick, Fournier Stephane, Bize Alain, Sambin Lucien, Berdeaux Alain, Varenne Olivier, De Bruyne Bernard, Ghaleh Bijan

Аннотация

Background Absolute hyperemic coronary blood flow and microvascular resistances can be measured by continuous thermodilution with a dedicated infusion catheter. We aimed to determine the mechanisms of this hyperemic response in animal. Methods and Results Twenty open chest pigs were instrumented with flow probes on coronary arteries. The following possible mechanisms of saline-induced hyperemia were explored compared with maximal hyperemia achieve with adenosine by testing: (1) various infusion rates; (2) various infusion content and temperature; (3) NO production inhibition with L-arginine methyl ester and endothelial denudation; (4) effects of vibrations generated by rotational atherectomy and of infusion through one end-hole versus side-holes. Saline infusion rates of 5, 10 and 15 mL/min did not reach maximal hyperemia as compared with adenosine. Percentage of coronary blood flow expressed in percent of the coronary blood flow after adenosine were 48±17% at baseline, 57±18% at 5 mL/min, 65±17% at 10 mL/min, 82±26% at 15 mL/min and 107±18% at 20 mL/min. Maximal hyperemia was observed during infusion of both saline at body temperature and glucose 5%, after endothelial denudation, l-arginine methyl ester administration, and after stent implantation. The activation of a Rota burr in the first millimeters of the epicardial artery also induced maximal hyperemia. Maximal hyperemia was achieved by infusion through lateral side-holes but not through an end-hole catheter. Conclusions Infusion of saline at 20 mL/min through a catheter with side holes in the first millimeters of the epicardial artery induces maximal hyperemia. The data indicate that this vasodilation is related neither to the composition/temperature of the indicator nor is it endothelial mediated. It is suggested that it could be elicited by epicardial wall vibrations.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты