Therapeutic drug monitoring · 2016
Monitoring Plasma Levels of Donepezil, 5-O-Desmethyl-Donepezil, 6-O-Desmethyl-Donepezil, and Donepezil-N-Oxide by a Novel HPLC Method in Patients With Alzheimer Disease.
СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Groppa Francesca, Coin Alessandra, De Rosa Giovanni, Granziera Serena, Alexopoulos Chiara, Pamio Maria Valentina, Padrini Roberto
Аннотация
BACKGROUND: In humans, donepezil (D) is metabolized to 5-O-desmethyl-donepezil (5DD), 6-O-desmethyl-donepezil (6DD), and donepezil-N-oxide (DNox). Although 6DD and DNox are pharmacologically active, the activity of 5DD is unknown. At present, no routine methods are available to detect D and its 3 metabolites simultaneously. In this study, a novel high-performance liquid chromatography method was developed and applied to a population of patients with Alzheimer disease on stable treatment with the drug.
METHODS: Liquid-liquid extraction from plasma was accomplished by means of a solvent mixture of n-hexane/dichloromethane/ethylacetate (45:40:15) after sample alkalinization. Disopyramide was the internal standard. After evaporation, the residue was reconstituted in 200 μL of mobile phase (acetonitrile 85%:1% acetic acid 15%) and 50 μL was injected into the high-performance liquid chromatography column (X-Terra, RP8; flow: 1 mL/min). Photometric and fluorimetric detectors were used in tandem, to maximize the sensitivity of fluorescent compounds (D, 5DD, and DNox) and also to reveal nonfluorescent compounds (6DD and internal standard).
RESULTS: The method was linear in the 10-100 ng/mL concentration range. Imprecision (coefficient of variation) varied between 3.2% and 12.6% and inaccuracy (% mean absolute error) between 1.3% and 13.3%, depending on the compound, concentration, and detection mode. The quantitation limits were 0.1-0.3 ng/mL for fluorescent compounds and 1.2-4.3 ng/mL for photometric compounds. D, 5DD, 6DD, and DNox through concentrations were measured in 54 patients with Alzheimer disease on treatment with D (10 mg q.d.). No interfering peaks by endogenous compounds or coadministered drugs were noted. Plasma concentrations were quite variable among patients (D: 10-106 ng/mL; 5DD: 0.07-2.8 ng/mL; 6DD: 1.2-36 ng/mL; DNox: 0.5-45.4 ng/mL). Of note, in 6 patients, the plasma concentrations of the 2 active metabolites (6DD and DNox) were higher than those of the parent drug.
CONCLUSIONS: The above method proved to be suitable for therapeutic drug monitoring and may be useful in ascertaining the real contribution of metabolites to the therapeutic effects of donepezil.
METHODS: Liquid-liquid extraction from plasma was accomplished by means of a solvent mixture of n-hexane/dichloromethane/ethylacetate (45:40:15) after sample alkalinization. Disopyramide was the internal standard. After evaporation, the residue was reconstituted in 200 μL of mobile phase (acetonitrile 85%:1% acetic acid 15%) and 50 μL was injected into the high-performance liquid chromatography column (X-Terra, RP8; flow: 1 mL/min). Photometric and fluorimetric detectors were used in tandem, to maximize the sensitivity of fluorescent compounds (D, 5DD, and DNox) and also to reveal nonfluorescent compounds (6DD and internal standard).
RESULTS: The method was linear in the 10-100 ng/mL concentration range. Imprecision (coefficient of variation) varied between 3.2% and 12.6% and inaccuracy (% mean absolute error) between 1.3% and 13.3%, depending on the compound, concentration, and detection mode. The quantitation limits were 0.1-0.3 ng/mL for fluorescent compounds and 1.2-4.3 ng/mL for photometric compounds. D, 5DD, 6DD, and DNox through concentrations were measured in 54 patients with Alzheimer disease on treatment with D (10 mg q.d.). No interfering peaks by endogenous compounds or coadministered drugs were noted. Plasma concentrations were quite variable among patients (D: 10-106 ng/mL; 5DD: 0.07-2.8 ng/mL; 6DD: 1.2-36 ng/mL; DNox: 0.5-45.4 ng/mL). Of note, in 6 patients, the plasma concentrations of the 2 active metabolites (6DD and DNox) were higher than those of the parent drug.
CONCLUSIONS: The above method proved to be suitable for therapeutic drug monitoring and may be useful in ascertaining the real contribution of metabolites to the therapeutic effects of donepezil.