INTRODUCTION

Our retrospective study aimed at assessing safety of vascular closure devices (VCDs) used in a large single-centre Interventional Radiology (IR) department. Complication and deployment failure rates using collagen-based (Angio-seal) and suture-based (ProGlide) devices for common femoral artery haemostasis were compared.

MATERIALS AND

METHODS

Data from VCDs deployed over a 6-year period were retrospectively analysed for patient age, procedure indication, puncture mode (antegrade/retrograde), sheath size, deployment failure and complications (haematoma, pseudoaneurysm formation, limb occlusion). Numerical and statistical analysis was undertaken.

RESULTS

Overall, 1321 common femoral artery punctures in 1217 patients were closed using VCDs. Failure rate using ProGlide was significantly higher when compared with Angio-seal (P=<0.001) in sheath sizes ≤8 Fr. Heparin was not administered in embolisation procedures compared with angioplasty with or without stenting. Therefore, haematoma tended to occur more frequently following angioplasty without stenting (P = 0.003) and angioplasty with stenting (P = 0.001), when compared with embolisation. Deployment failure occurred more frequently when heparin was used during the procedure (P = 0.005).

CONCLUSION

Although complications relating to sheath size are well established in the literature, there remains a paucity of data assessing the impact of procedure specific factors when comparing VCDs. Our study challenges that size is the sole determinant of VCD success and invites a more holistic view of VCD deployment strategies. This study advocates continued research into the nuances of other potential confounding variables to optimise patient outcomes.

Для цитирования: Sethi Sifut, Michalski Jakub, Moh'd Elayyan Al-Shboul Rand, Carey Frank, Tan Kelvin, Ali Tariq. Procedure-Related Complication Rates With the Use of Vascular Closure Devices; Does Size Only Matter? A Large Single Centre Retrospective Study.. Vascular and endovascular surgery. 2024. DOI: 10.1177/15385744241276688.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.