DOI 10.1097/phm.0b013e31816f17e1
Gait analysis on split-belt force treadmills: validation of an instrument.
Аннотация
OBJECTIVE
Gait analysis (GA) is usually performed during overground walking, with ground reactions being recorded through force platforms embedded in the floor. Body kinematics and surface electromyography (EMG) are also recorded. GA on treadmills resting on force sensors (GAFT) is rather uncommon and has been criticized on various grounds (less reliability of force records; biomechanical differences in walking kinematics and dynamics). On the other hand, the treadmill allows very fast recording of data from many successive strides, at known and constant average speeds. The goals of this study were (a) to validate a system for GAFT based on a commercially available "split-belt" force treadmill, (b) to provide evidence that data can be compared with established norms from overground GA, and (c) to lend support to the potential clinical validity of the GAFT approach.
DESIGN
The treadmill adopted here allows subjects to walk on two parallel independent treadmills. Each of them is mounted on four 3D force sensors. Eight healthy adults (four women; ages 22-35 yrs) were tested on the treadmill, each at the average walking speed he or she adopted overground (average across subjects: 1.35 +/- 0.05 m x sec(-1)). Ground reactions, hip, knee, and ankle sagittal rotations, torques, power, and surface EMG from four thigh and leg muscles were recorded simultaneously.
RESULTS
Results were compared with those relating to a sample of 40 healthy adults (20 women; manufacturer's data, replicating published data) walking on floor-embedded force platforms (age 20-40 yrs; speed 1.33 +/- 0.06 m x sec(-1)). An 8% shorter stride length was adopted on the treadmill. All of the other results matched those obtainable overground.
CONCLUSIONS
The results suggest that GAFT performed on the tested treadmill is a promising method of GA in a clinical setting.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.