Assessment of the geometric profile of the Amplatzer and Cardioseal septal occluders by three dimensional echocardiography.

Acar P, Saliba Z, Bonhoeffer P, Sidi D, Kachaner J

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To apply three dimensional echocardiography to describe the geometric profile of the Amplatzer and Cardioseal occluders after deployment for closure of atrial septal defect.

METHODS

20 patients (mean (SD) age, 14 (5) years) were enrolled for transcatheter closure of a secundum atrial septal defect with the Amplatzer occluder (10) or with the Cardioseal occluder (10). The two populations were matched for the stretched diameter of the defect (mean 18 (6) mm). The profile of the two occluders was examined.

RESULTS

Transoesophageal echocardiography did not show any residual shunts after Amplatzer occluder deployment, whereas three patients had a small residual leak after Cardioseal deployment. One patient had transient atrioventricular block with the Amplatzer device. The mean surface area of the Amplatzer occluder was 6.9 (2) cm(2), and that of the Cardioseal device 5.4 (3) cm(2) (p = 0.03). The mean volume of the Amplatzer occluder was 9.2 (1) cm(3), while that of the Cardioseal occluder was 3.5 (1) cm(3) (p < 0.0001). From the three dimensional views, the Cardioseal occluder looked like a flat square after deployment whereas the Amplatzer occluder took up a ball shape in the atrial cavity.

CONCLUSIONS

Three dimensional views by multiplane transoesophageal echocardiography allow a realistic in vivo description of atrial septal occluders. The Amplatzer occluder, with its high geometric profile, allows complete closure of large atrial septal defects but with some risk of mechanical complications. Use of the Cardioseal device, with its small surface coverage and high residual shunt rate, should be limited to transcatheter closure of a patent foramen ovale or small atrial septal defects.

Для цитирования: Acar P, Saliba Z, Bonhoeffer P, Sidi D, Kachaner J. Assessment of the geometric profile of the Amplatzer and Cardioseal septal occluders by three dimensional echocardiography.. Heart (British Cardiac Society). 2001. DOI: 10.1136/heart.85.4.451.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.