Right ventricular volume unloading evaluated by tangential magnetocardiography.
Аннотация
OBJECTIVE
The evaluation of right ventricular volume overload in the presence of a right bundle branch block based solely on the results of an electrocardiogram is difficult. The purpose of this study was to purify acute right ventricular volume unloading from the tangential magnetocardiography.
METHODS
We measured the tangential (x-y plane) magnetocardiogram and electrocardiogram simultaneously in nine patients with a secundum atrial septal defect. The magnetocardiograms were obtained before surgical closure and during the immediate postoperative period using the 32-channel superconducting quantum interference device system in a magnetically shielded room.
RESULTS
The QRS duration on the surface electrocardiogram decreased significantly (p < 0.05) in the immediate postoperative period. The right ventricular depolarization time as measured by the magnetocardiogram was shortened from 40.3 +/- 6.1 to 25.3 +/- 7.4 milliseconds (p < 0.0005). The maximum peak amplitude during right ventricular depolarization decreased from 17.9 +/- 4.8 to 11.1 +/- 4.9 pT (p < 0.01).
CONCLUSIONS
We conclude that acute volume unloading of the right ventricle was indicated quantitatively by shortening of the right ventricular depolarization time and a reduction in the amplitude of current vectors originating from the right ventricular depolarization on the tangential magnetocardiography.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.