MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s12149-010-0354-1

Медицинская библиотека · Annals of nuclear medicine · 2010

Basic performance and stability of a CdTe solid-state detector panel.

Tsuchiya Katsutoshi, Takahashi Isao, Kawaguchi Tsuneaki, Yokoi Kazuma, Morimoto Yuuichi, Ishitsu Takafumi, Suzuki Atsurou, Ueno Yuuichirou, Kobashi Keiji

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

We have developed a prototype gamma camera system (R1-M) using a cadmium telluride (CdTe) detector panel and evaluated the basic performance and the spectral stability.

METHOD: The CdTe panel consists of 5-mm-thick crystals. The field of view is 134 x 268 mm comprising 18,432 pixels with a pixel pitch of 1.4 mm. Replaceable small CdTe modules are mounted on to the circuit board by dedicated zero insertion force connectors. To make the readout circuit compact, the matrix read out is processed by dedicated ASICs. The panel is equipped with a cold-air cooling system. The temperature and humidity in the panel were kept at 20 degrees C and below 70% relative humidity. CdTe polarization was suppressed by the bias refresh technique to stabilize the detector. We also produced three dedicated square pixel-matched collimators: LEGP (20 mm-thick), LEHR (27 mm-thick), and LEUHR (35 mm-thick). We evaluated their basic performance (energy resolution, system resolution, and sensitivity) and the spectral stability in terms of short-term (several hours of continuous acquisition) and long-term (infrequent measurements over more than a year) activity.

RESULTS

The intrinsic energy resolution (FWHM) acquired with Tc-99m (140.5 keV) was 6.6%. The spatial resolutions (FWHM at a distance of 100 mm) with LEGP, LEHR, and LEUHR collimators were 5.7, 4.9, and 4.2 mm, and the sensitivities were 71, 39, and 23 cps/MBq, respectively. The energy peak position and the intrinsic energy resolution after several hours of operation were nearly the same as the values a few minutes after the system was powered on; the variation of the peak position was <0.2%, and that of the resolution was about 0.3%. Infrequent measurements conducted over a year showed that the variations of the energy peak position and the intrinsic energy resolution of the system were at a similar level to those described above.

CONCLUSION

The basic performance of the CdTe-gamma camera system was evaluated, and its stability was verified. It was shown that the camera could be operated daily for several months without calibration.

Для цитирования: Tsuchiya Katsutoshi, Takahashi Isao, Kawaguchi Tsuneaki, Yokoi Kazuma, Morimoto Yuuichi, Ishitsu Takafumi, Suzuki Atsurou, Ueno Yuuichirou, Kobashi Keiji. Basic performance and stability of a CdTe solid-state detector panel.. Annals of nuclear medicine. 2010. DOI: 10.1007/s12149-010-0354-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.