DOI 10.1007/s12149-009-0293-x

Медицинская библиотека · Annals of nuclear medicine · 2009

Development of an ultra-high resolution SPECT system with a CdTe semiconductor detector.

Ogawa Koichi, Ohmura Naoka, Iida Hirokazu, Nakamura Kayoko, Nakahara Tadaki, Kubo Atsushi

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

The aim of this work was to evaluate an ultra-high spatial resolution SPECT system with a semiconductor detector and a high-resolution parallel-hole collimator or a pinhole collimator for small animal imaging.

METHODS

We evaluated an ultra-high spatial resolution SPECT system with a high-resolution parallel-hole collimator attached to a cadmium telluride (CdTe) semiconductor detector for small animal imaging. The sizes of an active area and a pixel in the semiconductor detector were 44 x 44 and 0.5 x 0.5 mm(2), respectively. In the high-resolution parallel-hole collimator the size of a hole was 0.4 x 0.4 mm(2), the thickness of a septum 0.1 mm, and the hole-length 30 mm. We also used a high-resolution pinhole collimator with a hole size of 0.3 or 0.5 mmvarphi. The physical performance of this SPECT system was evaluated with some experiments with phantoms filled with (99m)Tc-pertechnatate solution. In addition ideal performance and limitations of the system were evaluated with Monte Carlo simulations under the same geometrical conditions as in the experiments. In the evaluation for small animal imaging, we used mice that were administered with (99m)Tc-MDP. We also conducted an ultra-high resolution X-ray CT of the mice to verify the accumulated location of (99m)Tc-MDP using the bone CT images of the mice.

RESULTS

The results of the phantom experiments showed that we could resolve 1 mmvarphi hot-channels and 1.6 mmvarphi cold-rods with the high-resolution parallel-hole collimator and pinhole collimators. We could image 0.3 mmvarphi hot-channels with the high-resolution pinhole collimators. The results of the simulations showed that the resolution limit in the pinhole imaging was about 0.6 mm FWHM. And the results of experiments with mice showed that we could reconstruct high-resolution images of (99m)Tc-MDP. Furthermore, the distribution of (99m)Tc-MDP in a mouse was found to correspond closely to the location of the bones of the mouse in reconstructions made with the ultra-high resolution X-ray CT system.

CONCLUSIONS

Our results demonstrated that the ultra-high spatial resolution SPECT system was feasible for small animal imaging allowing a relatively long data acquisition time.

Для цитирования: Ogawa Koichi, Ohmura Naoka, Iida Hirokazu, Nakamura Kayoko, Nakahara Tadaki, Kubo Atsushi. Development of an ultra-high resolution SPECT system with a CdTe semiconductor detector.. Annals of nuclear medicine. 2009. DOI: 10.1007/s12149-009-0293-x.