DOI 10.1002/rcm.8738

Медицинская библиотека · Rapid communications in mass spectrometry : RCM · 2020

Phosphine detection in veterinary samples using headspace gas chromatography/tandem mass spectrometry with multiple reaction monitoring.

Lehner Andreas F, Johnson Margaret, Buchweitz John P

СтатьяПеревод в очереди

Аннотация

RATIONALE: Determination of phosphine exposure from zinc or aluminum phosphide fumigants continues to be a routine analytical requirement in veterinary forensic toxicology. There is a need for a more reliable and specific method than simple gas chromatography/mass spectrometry (GC/MS) analysis of sample solvent extracts, as GC/MS of extracts on capillary columns used for general screens involves significant interference from air peaks.

METHODS

GC/MS/MS headspace analysis of acid-generated phosphine gas enabled study of the feasibility of devising multiple reaction monitoring (MRM) approaches to the determination of phosphine with greater specificity.

RESULTS

Collision-induced dissociation in GC/MS/MS showed that phosphine generated m/z 34 → 31, 32 and 33 ion transitions by sequential proton release as well as minor transitions m/z 34 → 47, 34 → 63 and 63 → 31.5 by intermolecular collisions and double charging. Study of the formation of these product ions enabled development of MRM settings for a highly useful headspace method for phosphine detection.

CONCLUSIONS

The method was validated over a working range of 5-100 ppm of phosphide generating phosphine gas which enabled retention of regular screen capillary columns without necessitating separation from air components. The method should have adequate sensitivity and reliability for veterinary toxicology laboratories confronting specimens from animals poisoned by crop fumigants.

Для цитирования: Lehner Andreas F, Johnson Margaret, Buchweitz John P. Phosphine detection in veterinary samples using headspace gas chromatography/tandem mass spectrometry with multiple reaction monitoring.. Rapid communications in mass spectrometry : RCM. 2020. DOI: 10.1002/rcm.8738.

Предметные рубрики MeSH