DOI 10.1111/cod.13924

Медицинская библиотека · Contact dermatitis · 2021

Contact dermatitis from glucose sensors in Spain: A multicentric approach.

Gatica-Ortega María-Elena, Mowitz Martin, Pastor-Nieto María-Antonia, Navarro-Triviño Francisco J, Fernández-Redondo Virginia, Hernández-Cano Natalia, Borrego Leopoldo, Gómez-de-la-Fuente Enrique, Torralba Miguel, Svedman Cecilia, Giménez-Arnau Ana

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

Allergic contact dermatitis from glucose sensors may interfere with their ongoing application.

OBJECTIVE

To evaluate a series of Spanish patients with contact dermatitis to glucose sensors regarding former sources of contact allergens, patch test results, and outcomes from the ongoing use of the device.

METHODS

A series of patients with contact dermatitis from glucose sensors was investigated in eight dermatology departments across Spain (epidemiological features, brands, latency time to develop dermatitis, the ability to continue using the devices as well as the patch test results).

RESULTS

Thirty patients were evaluated (mean age 20.9 years). A total of 66.7% were children and 66.7% female. Ninety per cent used Freestyle Libre (FSL). Eight of 26 (30.8%) reacted to isobornyl acrylate (IBOA) and two of 20 (10.0%) to N,N dimethylacrylamide (DMAA). The mean latency time to develop dermatitis was 9 months. Sixteen of 29 (55.2%) patients continued using the same sensor causing the reaction. Thirteen of 29 (44.8%) patients were unable to continue using the sensor because of severe reactions. Of these, five were positive to IBOA, one to IBOA and DMAA, one to DMAA, one to colophony, and one to isopropyl alcohol wipes. In one patient, the outcome was unknown.

CONCLUSION

The frequency of sensitisation to IBOA and DMAA, was lower than in other European series, but similar to a previously published Spanish article. Legislation requiring manufacturers to provide information regarding the composition of medical devices and cooperate with the investigations into contact dermatitis is urgently needed.

Для цитирования: Gatica-Ortega María-Elena, Mowitz Martin, Pastor-Nieto María-Antonia, Navarro-Triviño Francisco J, Fernández-Redondo Virginia, Hernández-Cano Natalia, Borrego Leopoldo, Gómez-de-la-Fuente Enrique, Torralba Miguel, Svedman Cecilia, Giménez-Arnau Ana. Contact dermatitis from glucose sensors in Spain: A multicentric approach.. Contact dermatitis. 2021. DOI: 10.1111/cod.13924.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.