PURPOSE: To analyze different control-system limitations on the measured dose distributions in low-dose regions of simplified intensity fields with an electronic portal imaging device to ascertain the optimal settings for the control-system limitations in the planning system.

MATERIAL AND

METHODS

The authors created one field with an "optimal fluence" of intensity 1.0 (full dose) and one field with intensity 0.0 (no dose) in the central part of the field. The influence of different dose rates (DRs) and maximum leaf speeds (LS) on the calculated and measured dose and dose profiles were analyzed.

RESULTS

Good agreement between calculated and measured dose in the case of a field of intensity 1.0 was found. For the field with intensity 0.0, the measured dose was 20-60% lower than the dose calculated by the "actual fluence". The results were found dependent on the DR and LS.

CONCLUSION

The overestimation in regions of optimal intensity 0.0 by the planning system cannot be resolved by the user. Taking the measured dose in the region of desired intensity 1.0 and other technical limitations (like beam hold interrupts or spikes in the cross and longitudinal profiles) into consideration, the application of an LS of 2.5 cm/s and a DR of 500 MU/min is recommended in order to minimize radiation dose applied to organs at risk, which are located in regions of low intensity, like, for example, the spinal cord.

Для цитирования: Vorwerk Hilke, Wagner Daniela, Seitz Björn, Christiansen Hans, Wolff Hendrik A, Hess Clemens F. Overestimation of low-dose radiation in intensity-modulated radiotherapy with sliding-window technique.. Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al]. 2009. DOI: 10.1007/s00066-009-2028-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.