Algorithm for Rapid Exclusion of Clinically Relevant Plasma Levels of Direct Oral Anticoagulants in Patients Using the DOAC Dipstick: An Expert Consensus Paper.
Аннотация
BACKGROUND
With the widespread use of direct oral anticoagulants (DOACs), there is an urgent need for a rapid assay to exclude clinically relevant plasma levels. Accurate and rapid determination of DOAC levels would guide medical decision-making to (1) determine the potential contribution of the DOAC to spontaneous or trauma-induced hemorrhage; (2) identify appropriate candidates for reversal, or (3) optimize the timing of urgent surgery or intervention.
METHODS AND
RESULTS
The DOAC Dipstick test uses a disposable strip to identify factor Xa- or thrombin inhibitors in a urine sample. Based on the results of a systematic literature search followed by an analysis of a simple pooling of five retrieved clinical studies, the test strip has a high sensitivity and an acceptably high negative predictive value when compared with levels measured with liquid chromatography tandem mass spectrometry or calibrated chromogenic assays to reliably exclude plasma DOAC concentrations ≥30 ng/mL.
CONCLUSION
Based on these data, a simple algorithm is proposed to enhance medical decision-making in acute care indications useful primarily in hospitals not having readily available quantitative tests and 24/7. This algorithm not only determines DOAC exposure but also differentiates between factor Xa and thrombin inhibitors to better guide clinical management.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.