DOI 10.1093/rpd/ncq366

Медицинская библиотека · Radiation protection dosimetry · 2010

Evaluation of the performance of two LiF:Mg,Ti and LiF:Mg,Cu,P dosemeters for extremity monitoring.

Freire L C, Pereira M F, Calado A M, Santos L M, Cardoso J V, Alves J G

СтатьяПеревод в очереди

Аннотация

In this paper, the results aimed at assessing the performance of two varieties of LiF detectors (LiF:Mg,Ti and LiF:Mg,Cu,P) in photon fields relatively to reproducibility, detection threshold and angular dependence as defined in the ISO 12794 standard are presented. The fading properties and the limit of detection were also investigated for both materials. The results suggest that both LiF varieties are well suited for extremity monitoring. However, better fading properties of LiF:Mg,Cu,P when compared with LiF:Mg,Ti, combined with previous results relatively to energy dependence suggests that LiF:Mg,Cu,P dosemeters are better suited for extremity monitoring.

Для цитирования: Freire L C, Pereira M F, Calado A M, Santos L M, Cardoso J V, Alves J G. Evaluation of the performance of two LiF:Mg,Ti and LiF:Mg,Cu,P dosemeters for extremity monitoring.. Radiation protection dosimetry. 2010. DOI: 10.1093/rpd/ncq366.