Implementation of at-line capillary zone electrophoresis for fast and reliable determination of adenovirus concentrations in vaccine manufacturing.
Аннотация
A CZE method was validated and implemented for fast and accurate in-process determination of adenovirus concentrations of downstream process samples obtained during manufacturing of adenovirus vector-based vaccines. An analytical-quality-by-design approach was embraced for method development, method implementation, and method maintenance. CZE provided separation of adenovirus particles from sample matrix components, such as cell debris, residual DNA and proteins. The intermediate precision of the virus particle concentration was 6.9% RSD and the relative bias was 2.3%. In comparison, the CZE method is intended to replace a quantitative polymerase chain reaction method which requires three replicates in three analytical runs to achieve an intermediate precision of 8.1% RSD. Given that, in addition, the time from sampling till reporting results of the CZE method was less than 2 h, whereas quantitative polymerase chain reaction requires 3 days, it follows that the CZE method enables faster processing times in downstream processing.