MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/bsd.0b013e31819942cd

Медицинская библиотека · Journal of spinal disorders & techniques · 2010

Pedicle screw augmentation with polyethylene tape: a biomechanical study in the osteoporotic thoracolumbar spine.

Hamasaki Takahiko, Tanaka Nobuhiro, Kim JinHwan, Okada Motohiro, Ochi Mitsuo, Hutton William C

СтатьяПеревод в очереди

Аннотация

STUDY

DESIGN

Biomechanical study using human thoracolumbar spines.

OBJECTIVE

To assess the strength of fixation of pedicle screws augmented with Nesplon tape applied sublaminar or applied subpars.

SUMMARY OF BACKGROUND DATA: There are no reports on the biomechanical assessment of Nesplon tape used for pedicle screw augmentation.

METHODS

Experiment 1: pedicle screws were inserted into bilateral pedicles as follows: (1) pedicle screw alone connected to a rod (PS alone), (2) pedicle screw augmented with Nesplon tape applied sublaminar, connected to a rod [pedicle screw applied sublaminar (PSSL)], and (3) pedicle screw augmented with tape applied subpars, connected to a rod [pedicle screw applied subpars (PSSP)]. The rods were pulled and pushed until the pedicle screw was loose. Experiment 2: 6 thoracolumbar spines were biomechanically tested as follows: axial compression, flexion, extension, left and right lateral bending, and left and right axial rotation. This sequence was applied to: (1) the intact spine; (2) the spine made unstable; (3) the spine fixed by pedicle screws and rods (PS alone); (4) the same as 3, but with the pedicle screws augmented using Nesplon tape applied sublaminar (PSSL); and (5) the same as 3, but with pedicle screws augmented using tape applied subpars (PSSP). From the load-deformation curves, stiffness values were calculated.

RESULTS

Experiment 1: the pedicle screws augmented by sublaminar or subpars tape (PSSL or PSSP) showed significantly greater fixation strength compared with those pedicles that were not augmented by tape (PS alone). Experiment 2: a pedicle screw/rod construct augmented by tape provided a stiffer construct than the same construct without augmentation.

CONCLUSIONS

Pedicle screws that are inserted into the osteoporotic thoracolumbar spine and augmented by Nesplon tape applied sublaminar or subpars provide firmer fixation of the screws and a stiffer pedicle screw/rod construct than the same construct without augmentation by tape.

Для цитирования: Hamasaki Takahiko, Tanaka Nobuhiro, Kim JinHwan, Okada Motohiro, Ochi Mitsuo, Hutton William C. Pedicle screw augmentation with polyethylene tape: a biomechanical study in the osteoporotic thoracolumbar spine.. Journal of spinal disorders & techniques. 2010. DOI: 10.1097/bsd.0b013e31819942cd.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.