MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.idh.2025.08.004

Медицинская библиотека · Infection, disease & health · 2025

Validation of performance qualification of ultrasound probe high-level disinfection devices in clinical settings.

Bellamy David, Vickery Karen

СтатьяВалидационное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

In accordance with AS5369:2023, disinfection processes need to undergo routine performance qualification (PQ) as part of in-field validation. Automated high-level disinfection (HLD) devices therefore need to undergo PQ. Responding to reports that some ultraviolet-C (UV-C) devices used for reprocessing of ultrasound transducers were failing to achieve sporicidal efficacy during routine PQ, we evaluated the sporicidal efficacy of these devices.

METHODS

Sporicidal efficacy was assessed using commercially available stainless steel biological indicators (BIs) inoculated with 106Geobacillus stearothermophilus spores (ATCC® 7953). BIs were clamped in top and bottom positions inside the chambers of the devices [UV-C light-emitting diode (LED), UV-C lamp and hydrogen peroxide (H2O2) mist]. All three devices were in clinical use. BI test conditions included packaged, unwrapped and non-flamed as well as unwrapped and flame sterilised on the clamped coupon end. Results were evaluated on a pass (no growth) or fail (growth) basis.

RESULTS

The UV-C LED device failed to inactivate spores in all tested positions and conditions (n = 18). The UV-C lamp device passed 2/6 of the unwrapped and flamed tests but failed all other tests (n = 12). The H2O2 mist device passed all tests, inactivating spores for all conditions and chamber positions (n = 18).

CONCLUSION

The H2O2 mist device was the only device in this study that passed all tests, demonstrating sporicidal efficacy. Both UV-C devices failed to demonstrate sporicidal efficacy in this PQ study suggesting that clinicians need to monitor UV-C devices more frequently to ensure continued efficacy and patient safety.

Для цитирования: Bellamy David, Vickery Karen. Validation of performance qualification of ultrasound probe high-level disinfection devices in clinical settings.. Infection, disease & health. 2025. DOI: 10.1016/j.idh.2025.08.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.