Acoustic Radiation Force Optical Coherence Elastography of the Crystalline Lens: Safety.

Zevallos-Delgado Christian, Mekonnen Taye Tolu, Duvvuri Chaitanya, Rohman Leana, Schumacher Justin, Singh Manmohan, Aglyamov Salavat R, Twa Michael D, Parel Jean-Marie, Scarcelli Giuliano, Manns Fabrice, Larin Kirill V

СтатьяПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

PURPOSE: To assess the safety of acoustic radiation force optical coherence elastography in the crystalline lens in situ.

METHODS

Acoustic radiation force (ARF) produced by an immersion single-element ultrasound transducer (nominal frequency = 3.5 MHz) was characterized using a needle hydrophone and used for optical coherence elastography (OCE) of the crystalline lens. Preamplified signals at 50, 100, 250, 500, 750, 1000, and 1250 mV peak amplitude were tested on ex vivo porcine eyes (n = 21). Three-dimensional optical coherence tomography (OCT) and confocal microscopy images were acquired before and after ARF exposure to each signal amplitude to determine damage.

RESULTS

The acoustic intensity of the ultrasound transducer at 100-mV preamplified peak amplitude input demonstrated a signal-to-noise ratio high enough for tracking elastic wave propagation in the lens and spatial-peak pulse-average (SPPA) intensity of 24.1 W/cm² and mechanical index (MI) of 0.46. The SPPA intensity was lower than the U.S. Food and Drug Administration (FDA) safety limit (28 W/cm2), but the MI was twice the safety limit (0.23). OCT structural and confocal microscopy images showed damage only at levels exceeding 1150 W/cm2 and 3.2 for SPPA intensity and MI, respectively.

CONCLUSIONS

OCT and confocal microscopy showed that, even when the intensity exceeded FDA recommendations (>100 mV), no noticeable damage was observed. Although a further reduction in acoustic intensity is necessary to meet FDA safety limits, ARF-based elastography shows promise for safe clinical translation in quantitatively characterizing lenticular biomechanical properties.

TRANSLATIONAL RELEVANCE: This work assessed the safety standards for acoustic radiation force to be used in human lens elastography according to the FDA safety limits.

Для цитирования: Zevallos-Delgado Christian, Mekonnen Taye Tolu, Duvvuri Chaitanya, Rohman Leana, Schumacher Justin, Singh Manmohan, Aglyamov Salavat R, Twa Michael D, Parel Jean-Marie, Scarcelli Giuliano, Manns Fabrice, Larin Kirill V. Acoustic Radiation Force Optical Coherence Elastography of the Crystalline Lens: Safety.. Translational vision science & technology. 2024. DOI: 10.1167/tvst.13.12.36.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.