Human factors engineering validation study for a novel 0.1-mg epinephrine auto-injector.

Edwards Evan, Kessler Catherine, Cherne Naomi, Dissinger Emily, Shames Adam

СтатьяПеревод в очереди

Аннотация

Background

Anaphylaxis in infants and young children is increasing. Historically, epinephrine auto-injectors (EAI) were not available in a dose or platform designed for patients who weighed <15 kg, and, therefore, 0.15-mg EAIs were prescribed for these patients. Results of ultrasound studies indicate that currently marketed 0.15-mg EAIs have needle lengths that may strike bone in ∼29 to 43% of patients who weigh <15 kg and result in possible intraosseous injection. An EAI with a more weight-appropriate 0.1-mg dose and shorter needle length has been developed to potentially help minimize the risk of striking bone during epinephrine injection in patients who weigh 7.5-15 kg.

Objective

A human factors usability study was completed to validate the 0.1-mg EAI user interface.

Methods

This study was conducted with parents of children who were severely allergic so to evaluate simulated use of the 0.1-mg EAI by the intended user group. Fifteen participants were enrolled and received training on using the 0.1-mg EAI. Approximately 24 hours later, the participants completed a simulated emergency-use scenario by using the 0.1-mg EAI with an infant manikin. The primary end point was successful simulated administration of a meaningful epinephrine dose.

Results

All the participants simulated administration of a meaningful epinephrine dose. Fourteen participants successfully used the 0.1-mg EAI per the instructions for use in the simulated emergency-use scenario. One participant did not press the EAI against the thigh for the length of time defined for this critical task; however, the EAI was pressed long enough for complete delivery of the 0.1-mg dose.

Conclusion

This study validated the user interface of the 0.1-mg EAI for the intended users, uses, and use environments.

Для цитирования: Edwards Evan, Kessler Catherine, Cherne Naomi, Dissinger Emily, Shames Adam. Human factors engineering validation study for a novel 0.1-mg epinephrine auto-injector.. Allergy and asthma proceedings. 2018. DOI: 10.2500/aap.2018.39.4171.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.