DOI 10.1007/s10334-026-01330-1
Design and chemical composition of a reference phantom for 13C metabolic MRSI.
Аннотация
OBJECT: Development of a reproducible and long-term stable reference phantom for 13C metabolic MRSI, addressing the need for accessible calibration and quality assurance tools across sites and platforms.
MATERIALS AND
METHODS
The phantom was built from readily available labware (50 mL centrifuge tubes, NMR glass vials) and 3D-printed holders, designed to ensure sufficient coil loading and excellent B0 homogeneity. 13C-labeled reference solutions were stabilized by argon degassing, sodium azide addition, and airtight flame sealing, with optional gadolinium doping to tune T1/T2 relaxation. Performance was tested at 1.4 T, 3 T, and 11.7 T using 1H MRI, 13C spectroscopy, and 13C MRSI.
RESULTS
The reference solutions remained chemically and spectrally stable for up to 20 months under cooled, light-protected storage. Ethyl acetate provided a stable alternative to pyruvate, while gadolinium enabled the control of relaxation properties. Spectral assessment confirmed narrow linewidths, strong SNR, and reproducible peak separation across field strengths, validating the phantom's suitability for calibration and method development.
DISCUSSION: This low-cost, open-source design provides a durable and reproducible 13C phantom that complements existing dynamic and metabolite-specific models. It enables long-term standardization, supports multi-center studies, and facilitates robust quality assurance in hyperpolarized and thermal 13C MRSI.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.