Reliability of the NeuroTrend sensor system under hyperbaric conditions.

Henze Dirk, Bomplitz Michael, Radke Joachim, Clausen Tobias

Сравнительное исследованиеСтатьяВалидационное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

The goal of this study was to investigate the reliability of the multi-parameter sensor NeuroTrend in a hyperbaric environment for up to 3bar absolute pressure. Measurement of brain tissue oxygenation (ptiO2) under hyperbaric conditions is supposed to elucidate whether hyperbaric oxygenation therapy has the potential to improve ptiO2 to a clinically significant degree in pathological altered brain tissue after traumatic brain injury.

METHODS

The NeuroTrend sensor hose, filled with equilibrated plasma samples, was stored in a decompression chamber. The plasma samples were equilibrated with three different gas mixtures. After determination of the initial values for temperature, oxygen partial pressure (pO2), carbon dioxide partial pressure (pCO2) and hydrogen ion concentration (pH) in the plasma, the ambient pressure was stepwise increased from 0.1 to 3 bar. The same set-up was performed without increasing the ambient pressure.

RESULTS

No significant difference in the mean values for all 23 measurement points and for all parameters (pO2, pCO2, pH) of all 10 NeuroTrend sensors was found, under both normobaric and hyperbaric conditions.

CONCLUSION

The study demonstrated that an absolute ambient pressure up to 3 bar did not influence the measuring properties and the reliability of the NeuroTrend sensor.

Для цитирования: Henze Dirk, Bomplitz Michael, Radke Joachim, Clausen Tobias. Reliability of the NeuroTrend sensor system under hyperbaric conditions.. Journal of neuroscience methods. 2004. DOI: 10.1016/j.jneumeth.2003.09.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.