m+Medelevator.RU
← Поиск публикаций
Aesthetic surgery journal · 2021

Vibration Exposure Safety Guidelines for Surgeons Using Power-Assisted Liposuction (PAL).

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Wu Chaowen, Laswell Stephen, Mentz James A, Morales Rolando

Аннотация

BACKGROUND: As power-assisted liposuction (PAL) gains in popularity, plastic surgeons operating these devices experience occupational exposure to hand-transmitted vibration, which can result in hand-arm vibration syndrome, a debilitating neurovasculopathy.

OBJECTIVES: The objective of the study was to determine vibration exposure from the utilization of a PAL device during surgery to generate recommendations for safe use.

METHODS: Vibration emission of a commonly utilized PAL system (MicroAire-650, Surgical Instruments, Charlottesville, VA) was examined employing a vibration data logger under both controlled laboratory conditions and during 13 typical liposuction cases. Data were analyzed and compared with established safety limits of vibration exposure.

RESULTS: The experiments demonstrated a mean vibration magnitude of typical liposuction surgeries to be 5.69 ± 0.77 m/s2 (range, 4.59-6.27 m/s2), which is significantly higher than the manufacturer declared value of 3.77 m/s2. Cannula size was shown to be the most significant contributor to vibration magnitude, with larger cannulas causing more vibration transmission.

CONCLUSIONS: These results indicate that recommendations must be made to prevent undue occupational exposure to vibration from PAL. The MicroAire-650 can generally be safely utilized for less than 1.5 h/d. At exposure levels >1.5 h/d, there is increased risk of developing vibration-related injuries, and vibration-reducing strategies should be implemented. At exposure levels >6 h/d, the safety limit is exceeded and there is significantly increased risk of developing hand-arm vibration syndrome and vibration exposure should be halted.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты