Comparison of a collagen plug arterial closure device with manual compression after endovascular interventions for peripheral artery disease.

Fokkema Talje M, Minnee Robert C, Kock Gert-Aldert H, Blomjous Joost G A M, Vahl Anco C, Leijdekkers Vanessa J

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

Arterial closure devices (ACDs) seem to be safe and effective to obtain hemostasis for cardiac interventions and diagnostic vascular procedures. However, only limited evidence is presented on therapeutic interventions in patients with peripheral atherosclerotic disease. We compared ischemic and hemorrhagic complications of an ACD vs manual compression (MC) after puncture of the femoral artery for endovascular lower extremity procedures.

METHODS

The records of all subsequent endovascular interventions for peripheral artery disease performed in a large teaching hospital, from January 2012 to April 2014, were retrospectively reviewed. The primary outcome was an ischemic complication requiring reintervention. Secondary outcomes were hematoma (needing conservative or invasive intervention) and infection (needing intervention or antibiotics). Multivariate analysis was performed to identify predictors for any complication.

RESULTS

The Angio-Seal (St. Jude Medical, Minnetonka, Minn) closure device was used in 244 of 463 endovascular interventions (54%). Five patients (2%) who received an ACD needed emergency surgery for ischemic complications vs none in the MC group (P = .06). Hematomas occurred in 22 patients (9%) after ACD vs in six (3%) after MC (adjusted odds ratio [OR], 3.4; 95% confidence interval [CI], 1.4-8.9; P = .015). There were no infections in either group. ACD (OR, 3.8; 95% CI, 1.5-9.7) and current smoking (OR, 2.6; 95% CI, 1.01-6.7) were independent predictors for any complication.

CONCLUSIONS

This study demonstrates that ACD use in patients with peripheral artery disease can lead to serious adverse events resulting in increased morbidity. Therefore, the potential benefits of an ACD over MC should be carefully weighed.

Для цитирования: Fokkema Talje M, Minnee Robert C, Kock Gert-Aldert H, Blomjous Joost G A M, Vahl Anco C, Leijdekkers Vanessa J. Comparison of a collagen plug arterial closure device with manual compression after endovascular interventions for peripheral artery disease.. Journal of vascular surgery. 2016. DOI: 10.1016/j.jvs.2016.02.025.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.