Automatic atrial threshold measurement and adjustment in pediatric patients.

Hiippala Anita, Serwer Gerald A, Clausson Eva, Davenport Lynn, Brand Trina, Happonen Juha-Matti

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Automatic threshold measurement and output adjustment are used as default settings in modern pacemakers. The purpose of the study was to assess Atrial Capture Management (ACM) of Medtronic pacemakers in pediatric patients.

METHODS

Forty children were enrolled in two centers. Median age was 9.8 years (range 0.8-17.5 years). Half had undergone surgery for congenital heart defects; 45% of patients had an epicardial atrial lead. The pacing indication was atrioventricular block in 82% of patients and sinus node disease in 18%. Manually determined atrial thresholds and ACM measurements were compared.

RESULTS

ACM measurements were within the expected variation in 37/40 (93%) of the patients. In one patient the threshold was 0.625-V lower manually than with ACM. One patient had too high an intrinsic atrial rate for ACM to be able to measure threshold. The mean threshold at 0.4 ms was 0.69 +/- 0.32 V manually and 0.68 +/- 0.35 V with ACM (two-tailed paired t-test, P = 0.52) in all patients. The mean difference was 0.012 V (95% confidence interval: -0.027, 0.053). The mean endocardial threshold was 0.70 +/- 0.36 V manually and 0.69 +/- 0.38 V with ACM; epicardial threshold was 0.67 +/- 0.27 V manually and 0.68 +/- 0.32 V with ACM. The difference between the measurements was 0.012 V for endocardial and 0.014 V for epicardial leads. No atrial arrhythmias due to ACM measurements were observed.

CONCLUSIONS

ACM measures atrial thresholds reliably in pediatric patients with both endocardial and epicardial leads, allowing its use in both. Constant high intrinsic atrial rate may prevent automatic threshold measurement in young children.

Для цитирования: Hiippala Anita, Serwer Gerald A, Clausson Eva, Davenport Lynn, Brand Trina, Happonen Juha-Matti. Automatic atrial threshold measurement and adjustment in pediatric patients.. Pacing and clinical electrophysiology : PACE. 2009. DOI: 10.1111/j.1540-8159.2009.02619.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.