Fitting prelingually deafened adult cochlear implant users based on electrode discrimination performance.

Debruyne Joke A, Francart Tom, Janssen A Miranda L, Douma Kim, Brokx Jan P L

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

This study investigated the hypotheses that (1) prelingually deafened CI users do not have perfect electrode discrimination ability and (2) the deactivation of non-discriminable electrodes can improve auditory performance.

DESIGN

Electrode discrimination difference limens were determined for all electrodes of the array. The subjects' basic map was subsequently compared to an experimental map, which contained only discriminable electrodes, with respect to speech understanding in quiet and in noise, listening effort, spectral ripple discrimination and subjective appreciation.

STUDY SAMPLE: Subjects were six prelingually deafened, late implanted adults using the Nucleus cochlear implant.

RESULTS

Electrode discrimination difference limens across all subjects and electrodes ranged from 0.5 to 7.125, with significantly larger limens for basal electrodes. No significant differences were found between the basic map and the experimental map on auditory tests. Subjective appreciation was found to be significantly poorer for the experimental map.

CONCLUSIONS

Prelingually deafened CI users were unable to discriminate between all adjacent electrodes. There was no difference in auditory performance between the basic and experimental map. Potential factors contributing to the absence of improvement with the experimental map include the reduced number of maxima, incomplete adaptation to the new frequency allocation, and the mainly basal location of deactivated electrodes.

Для цитирования: Debruyne Joke A, Francart Tom, Janssen A Miranda L, Douma Kim, Brokx Jan P L. Fitting prelingually deafened adult cochlear implant users based on electrode discrimination performance.. International journal of audiology. 2016. DOI: 10.1080/14992027.2016.1243262.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.