DOI 10.1016/j.gaitpost.2022.05.007
Quantifying alignment bias during the fabrication and fitting of ankle-foot orthoses: A single center study.
Аннотация
BACKGROUND
The sagittal plane alignment of ankle-foot orthoses (AFO) and AFO footwear combinations (AFO-FC) has been shown to influence gait outcomes. As such, clinicians often target a particular alignment during the fabricating and fitting of an AFO to maximize outcomes.
RESEARCH QUESTION: How does the alignment of an AFO change during the fabrication and fitting process with respect to the intended, benchmark sagittal plane alignment identified by the consulting orthotist?
STUDY
DESIGN
Prospective
METHODS
The assessment of AFO alignment was performed using a convenience sample of 125 custom molded AFOs from 68 individuals fabricated at our center (57 bilateral AFOs, 11 unilateral AFOs). The alignment of each AFO was measured at 5 distinct steps during the fabrication and fitting process using a recently validated method to measure AFO neutral angle using differential inclinometers.
RESULTS
Prior to fabrication, the intended, benchmark alignment set by the consulting orthotist was 90 degrees for 92% of AFOs, was between 1 and 7 degrees of dorsiflexion for 7% of AFOs and was 5 degrees of plantarflexion for 1% of AFOs. Repeated measures ANOVA showed that AFO alignment changed between all fabrication and fitting steps. Overall, paired t-tests confirmed that AFO alignment was consistently 2-5 degrees more dorsiflexed than the benchmark alignment. Prior to fitting shoes, 55% of fabricated AFOs measured more than 2 degrees from the benchmark alignment. After fitting shoes, nearly 87% of AFO-FCs were more than 2 degrees from the benchmark alignment.
SIGNIFICANCE: The finding of systematic dorsiflexion bias and changes in AFO alignment throughout the fabrication and fitting process indicates the need to improve AFO fabrication precision. The neutral angle measurement methodology - using differential inclinometers - provides a means to improve this precision by enabling orthotists to precisely quantify and make appropriate adjustments to AFO alignment throughout the entire fabrication and fitting process.