DOI 10.1007/s00405-020-06031-6
Three-year audiological outcomes of the latest generation middle ear transducer (MET) implant.
Аннотация
PURPOSE: To evaluate the long-term audiological outcomes and safety of the latest generation of middle ear transducer (MET) among a group of Polish patients.
METHODS
Ten patients aged 48-72 years with bilateral sensorineural hearing loss (n = 8) and mixed hearing loss (n = 2) were included in this study. Pure tone audiometry, sound thresholds, word recognition scores in quiet and speech reception thresholds in noise were assessed. Medical and technical complication information was gathered.
RESULTS
All the patients underwent unilateral implantation with the latest generation Cochlear MET between 2014 and 2016. Mean length of follow-up was 3.7 years. Compared to the unaided condition, the implant provided significant functional gain (mean M = 26.1 dB) at 12 months follow-up. Compared to before surgery, average word recognition in quiet at 65 dB and at 80 dB SPL, as well as speech reception threshold in noise, were significantly better at 12 months. However, postoperative air conduction thresholds 6 months after implantation worsened by 10.3 dB (standard deviation SD = 5.8 dB). Postoperatively, three patients had skin problems around the processor, and one of them completely resigned from using the device 5 months after activation. Technical failures occurred in 4 cases. There were 9 out of 10 patients who still used the MET, but only 5 of them used the processor regularly (every day).
CONCLUSION
Despite changes in the transducer implemented by the manufacturer, we observed a significant number of adverse events in users of the latest generation of MET.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.