Comparison of subdural grid and stereoelectroencephalography in a cohort of pediatric patients.

Talai Afsaneh, Eschbach Krista, Stence Nicholas V, Palmer Claire, Bird Wynne, Alexander Allyson, O'Neill Brent R, Chapman Kevin E, Joshi Charuta

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare adverse events and outcomes between stereoelectroencephalography (SEEG) and subdural electrode (SDE) implantation in children.

METHODS

This was a retrospective analysis of 108 patients who underwent intracranial monitoring with SEEG or SDE implantation at Children's Hospital Colorado between January 2011 and June 2019.

RESULTS

There were 47 patients who underwent 53 SEEG implantations and 61 patients who underwent 64 SDE implantations, with an average age of 12.45 years (range: 1.22-19.96 years). Post-implantation imaging was performed in all SEEG implantations and 42 SDE implantations. 38 % and 88 % of SEEG and SDE implantations, respectively, had a hemorrhage of any kind (p < 0.01). Clinically significant hemorrhages did not differ between the two groups, though one death was reported in the SEEG group. No patient undergoing SEEG implantation received blood products compared to 20 % of SDE patients (p < 0.01). The rate of infection in SEEG patients was 4% compared to 33 % for SDE patients (p = 0.01). Resection was completed in 60 % of SEEG patients versus 93 % for SDE patients (p < 0.01). Rate of seizure response was not significantly different between the two groups, with 81 % and 71 % of SEEG and SDE patients, respectively, reaching Engel class I or II at 12 months (p = 0.76).

SIGNIFICANCE: In pediatric patients at a single institution, SEEG is associated with less adverse effects overall yet similar rates of seizure freedom compared to SDE implantation. This includes significantly lower rates of asymptomatic hemorrhage, infection and need for blood transfusion associated with SEEG monitoring. There was no statistical difference in clinically significant hemorrhages between the two groups, although rare in both.

Для цитирования: Talai Afsaneh, Eschbach Krista, Stence Nicholas V, Palmer Claire, Bird Wynne, Alexander Allyson, O'Neill Brent R, Chapman Kevin E, Joshi Charuta. Comparison of subdural grid and stereoelectroencephalography in a cohort of pediatric patients.. Epilepsy research. 2021. DOI: 10.1016/j.eplepsyres.2021.106758.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.