Comparison of bioimpedance versus thermodilution cardiac output during cardiac surgery: evaluation of a second-generation bioimpedance device.

Spiess B D, Patel M A, Soltow L O, Wright I H

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To compare a second-generation thoracic electrical bioimpedance (TEB) hemodynamic monitoring system with the clinically used pulmonary artery catheter thermodilution (TD-PAC) system.

DESIGN

Blinded, simultaneous measurements at specified key time points during surgery.

SETTING

University teaching hospital cardiac surgical operating rooms.

PARTICIPANTS

Forty-seven patients undergoing primary elective coronary artery bypass surgery.

INTERVENTIONS: Timed cardiac output measurements by thermodilution and continuous monitoring of bioimpedance were performed.

MEASUREMENTS AND MAIN

RESULTS

Cardiac index (TEB and TD-PAC) and other hemodynamic parameters were measured at 4 time points: (1) after anesthesia induction, (2) with the mediastinum open, (3) immediately after cardiopulmonary bypass, and (4) at the end of the case. Pearson's correlation and Bland-Altman analysis were carried out. Cardiac index by TEB and TD-PAC had an overall correlation of r = 0.71 (p < 0.0001). The Bland-Altman statistics showed a mean difference of -0.28 L/min/m2 and precision of 0.67 L/min/m2. The best correlation was at time 1, and the lowest correlation was at time 4. Mediastinal opening and cardiopulmonary bypass had little or no effect on the correlation between technologies.

CONCLUSION

TEB reporting of cardiac index during coronary artery surgery generally agreed with TD-PAC cardiac index except at the end of the case (time 4).

Для цитирования: Spiess B D, Patel M A, Soltow L O, Wright I H. Comparison of bioimpedance versus thermodilution cardiac output during cardiac surgery: evaluation of a second-generation bioimpedance device.. Journal of cardiothoracic and vascular anesthesia. 2001. DOI: 10.1053/jcan.2001.26533.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.