Tacrolimus Monitoring in Liver Transplant Recipients, Posttransplant Cholestasis: A Comparative Between 2 Commercial Immunoassays and a Liquid Chromatography-Tandem Mass Spectrometry Method.

Parant François, Delignette Marie-Charlotte, Charpiat Bruno, Lacaille Louis, Lebosse Fanny, Monneret Guillaume, Mohkam Kayvan, Mabrut Jean-Yves, Aubrun Frederic, Heyer Laurent, Antonini Teresa

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Cholestasis commonly occurs after orthotopic liver transplantation. It can be extrahepatic because of mechanical obstruction or intrahepatic because of various causes. During cholestasis episodes, blood concentrations of tacrolimus (TAC) metabolites may increase, potentially affecting TAC concentrations measured by immunoassays. This study aimed to simultaneously evaluate the analytical performance of 2 TAC immunoassays, a quantitative microsphere system (QMS) immunoassay, and chemiluminescence microparticle immunoassay, using liquid chromatography-tandem mass spectrometry (LC-MS/MS) as a reference method in liver transplant recipients.

METHODS

This single-center study included 265 patients who underwent orthotopic liver transplantation. In total, 942 blood samples were collected. TAC trough concentrations were measured using LC-MS/MS and 2 immunoassays in parallel. The plasma concentrations of conjugated bilirubin were measured in all samples. The results were analyzed using Bland-Altman plots and Passing-Bablok regressions.

RESULTS

The Bland-Altman plot analysis showed that the TAC QMS immunoassay has a significant bias (+37%) compared with LC-MS/MS, and this bias was higher in patients with cholestasis with hyperbilirubinemia (≤+70% in patients with conjugated bilirubin >150 µmol/L). In comparison, the chemiluminescence microparticle immunoassay showed acceptable analytical performance in patients with hyperbilirubinemia (bias <10%).

CONCLUSIONS

In agreement with previous findings, the TAC QMS immunoassay showed a positive bias compared with LC-MS/MS. This bias is remarkably high in patients with cholestasis and hyperbilirubinemia, suggesting the cross-reactivity of TAC metabolites with the monoclonal antibody used in the QMS immunoassay.

Для цитирования: Parant François, Delignette Marie-Charlotte, Charpiat Bruno, Lacaille Louis, Lebosse Fanny, Monneret Guillaume, Mohkam Kayvan, Mabrut Jean-Yves, Aubrun Frederic, Heyer Laurent, Antonini Teresa. Tacrolimus Monitoring in Liver Transplant Recipients, Posttransplant Cholestasis: A Comparative Between 2 Commercial Immunoassays and a Liquid Chromatography-Tandem Mass Spectrometry Method.. Therapeutic drug monitoring. 2024. DOI: 10.1097/ftd.0000000000001201.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.