SNR-efficient whole-brain pseudo-continuous arterial spin labeling perfusion imaging at 7 T.
Аннотация
PURPOSE: To optimize pseudo-continuous arterial spin labeling (PCASL) parameters to maximize SNR efficiency for RF power constrained whole brain perfusion imaging at 7 T.
METHODS
We used Bloch simulations of pulsatile laminar flow to optimize the PCASL parameters for maximum SNR efficiency, balancing labeling efficiency and total RF power. The optimization included adjusting the inter-RF pulse spacing (TRPCASL), mean B1 + (B1 + ave), slice-selective gradient amplitude (Gmax), and mean gradient amplitude (Gave). In vivo data were acquired from six volunteers at 7 T to validate the optimized parameters. Dynamic B0-shimming and flip angle adjustments were used to avoid needing to make the PCASL parameters robust to B0/B1 + variations.
RESULTS
The optimized PCASL parameters achieved a significant (3.3×) reduction in RF power while maintaining high labeling efficiency. This allowed for longer label durations and minimized deadtime, resulting in a 118% improvement in SNR efficiency in vivo compared to a previously proposed protocol. Additionally, the static tissue response was improved, reducing the required distance between labeling plane and imaging volume.
CONCLUSION
These optimized PCASL parameters provide a robust and efficient approach for whole brain perfusion imaging at 7 T, with significant improvements in SNR efficiency and reduced specific absorption rate burden.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.