MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10096-024-04822-x

Медицинская библиотека · European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology · 2024

Detection of intrathecal IgG antibody for varicella and measles diagnosis by evaluation and comparison of a commercial IgG chemiluminescent immunoassay with two ELISAs.

Garcia Rafael, Jiménez-Valera Maria, Ruiz-Buck Daniel, Sanchez Carlos, Rojas Almudena, Schütz Malte Hendrik, Rojas Jose, Hunfeld Klaus- Peter

СтатьяСравнительное исследованиеОценочное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

Analyse alternative methods of intrathecal antibody detection by comparing chemiluminescent immunoassay (CLIA) and enzyme-linked immunosorbent assay (ELISA) techniques to determine if CLIA can replace ELISA in the diagnosis of CNS infections.

METHODS

A panel of 280 paired samples-cerebrospinal fluid (CSF) and serum-with known antibody reactivities (Varicella, n = 60; Measles, n = 120) and negative samples (n = 100) were used to evaluate the performance of six serological test kits (Enzygnost, VirClia®, and Serion ELISA (Measles and Variella).

RESULTS

For Measles virus IgG, the VirClia® IgG monotest revealed 97% and 94% positive and negative agreement to the Enzygnost as reference test, respectively. In contrast, Serion ELISA kits yielded values of 18% and 90%. For the Varicella Zoster virus (VZV) IgG, the VirClia® IgG monotest showed 97% and 90% positive and negative agreement compared to Enzygnost. The Serion ELISA kits showed values of 55% and 86%, respectively. ROC analysis revealed that the areas under the curve for Measles and VZV IgGs were 0.7 and 0.852, respectively, using the Serion kit, and 0.963 and 0.955, for Vircell S.L CLIA technique. VirClia® monotest values were calculated using an antibody index cut-off of 1.3.

CONCLUSION

The findings indicate that CLIA testing can improve antibody detection in CSF samples, aiding the diagnosis of infectious neurological impairments.

Для цитирования: Garcia Rafael, Jiménez-Valera Maria, Ruiz-Buck Daniel, Sanchez Carlos, Rojas Almudena, Schütz Malte Hendrik, Rojas Jose, Hunfeld Klaus- Peter. Detection of intrathecal IgG antibody for varicella and measles diagnosis by evaluation and comparison of a commercial IgG chemiluminescent immunoassay with two ELISAs.. European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology. 2024. DOI: 10.1007/s10096-024-04822-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.