MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3233/thc-220091

Медицинская библиотека · Technology and health care : official journal of the European Society for Engineering and Medicine · 2023

A hip disarticulation prostheses test system to simulate gait for prostheses evaluation.

He Chen, Xiao Yixuan, Li Xinwei, Deng Zhipeng, Qian Yu, Zhao Jing, Yu Hongliu

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The limited number of hip prostheses users makes it less feasible to conduct amputee tests for prosthesis development in the clinic, which restricts the development efficiency of the intelligent prostheses.

OBJECTIVE

This study proposes a hip disarticulation prostheses test system (HDPTS) to supplement the amputee tests for hip disarticulation prosthesis (HDP) evaluation, which would potentially facilitate the prosthesis evaluation safety and development efficiency.

METHODS

The hip trajectory of an individual with normal gait was acquired and reproduced by calculating the corresponding movement joint angle of a manipulator. Then, an HDP was fit on an amputee and on the HDPTS respectively to obtain the hip and knee joint angles of the HDP during walking. Comparing the root mean square error (RMSE) of the expected and planned trajectory, the joint angles between the amputee test and HDPTS test, to verify the feasibility and accuracy of the HDPTS for prosthesis evaluation.

RESULTS

The RMSE between the expected and planned trajectory value was less than 1.20 mm (< 0.19%). The RMSE of the joint angles between the amputee test and HDPTS test were 2.18∘ (1.8%) and 3.13∘ (5.92%) for hip and knee joint respectively.

CONCLUSION

The HDPTS was found accurate in hip trajectory reproduction and feasible in gait simulation for the prosthesis evaluation, which could potentially supplement the amputee test for prosthesis design thus improving prosthesis test safety and development efficiency.

Для цитирования: He Chen, Xiao Yixuan, Li Xinwei, Deng Zhipeng, Qian Yu, Zhao Jing, Yu Hongliu. A hip disarticulation prostheses test system to simulate gait for prostheses evaluation.. Technology and health care : official journal of the European Society for Engineering and Medicine. 2023. DOI: 10.3233/thc-220091.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.