Idiopathic Lead Migration: Concept and Variants of an Uncommon Cause of Cardiac Implantable Electronic Device Dysfunction.

Morales José L, Nava Santiago, Márquez Manlio F, González Jorge, Gómez-Flores Jorge, Colín Luis, Martínez-Ríos Marco A, Iturralde Pedro

Клинический случайСтатьяОбзорПеревод пока не подготовлен

Аннотация

OBJECTIVES

This cumulative case study was performed to properly address the possible mechanisms, forms, and consequences of "twiddler's," "reel," and "ratchet" syndromes.

BACKGROUND

Twiddler's, reel, and ratchet syndromes are rare entities responsible for lead displacement of cardiac implantable electronic devices (CIED).

METHODS

From 2007 to 2012, 1,472 CIED were implanted at our center. Eighty-nine cases were reviewed for failure of pacing circuit integrity. Only 9 met the inclusion criteria for idiopathic lead migration (ILM) and were grouped as ILM (twiddler) or ILM (reel). For a pooled analysis of cases, a review of the literature from 1990 to 2012 was performed, and the authors identified 78 cases from 64 publications.

RESULTS

The study population consisted of 87 cases (45 women; median age, 66 years; 46 with ILM [twiddler] and 41 with ILM [reel]). Migration affected only 1 lead in 65% of 46 devices with more than 1 lead. None of the previously reported risk factors-manual manipulation of the device, elderly age, obesity, oversized pocket, and psychiatric history-correlated with the risk of ILM.

CONCLUSIONS

Neither manual manipulation of the device nor the other traditional risk factors reported in the literature for ILM syndrome correlated with the risk of ILM.

Для цитирования: Morales José L, Nava Santiago, Márquez Manlio F, González Jorge, Gómez-Flores Jorge, Colín Luis, Martínez-Ríos Marco A, Iturralde Pedro. Idiopathic Lead Migration: Concept and Variants of an Uncommon Cause of Cardiac Implantable Electronic Device Dysfunction.. JACC. Clinical electrophysiology. 2017. DOI: 10.1016/j.jacep.2017.02.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.