Paediatric cardiac catheterization radiation doses: a 6-year retrospective survey and typical values in a Mexican tertiary care center.
Аннотация
BACKGROUND
Establishing typical values is vital for radiation safety in paediatric fluoroscopically guided cardiac procedures (FGCP).
AIM: To publish the first typical values for FGCP in a Mexican tertiary hospital.
METHODS
We retrospectively analyzed 136 procedures (2018-2024), stratified by procedure type, weight, and age. Kerma-area product (Pka) and cumulative air kerma (Ka,r) were collected.
RESULTS
Typical values (median Pka) ranged from: <5 kg (2.84 Gy·cm2) up to 50-<80 kg (30.54 Gy·cm2). By age group <1 year (3.28 Gy·cm2) up to 10-15 years (20.63 Gy·cm2). No statistical difference was found between diagnostic and therapeutic procedures.
CONCLUSIONS
Weight-stratified institutional typical values revealed critical disparities, notably a five-fold higher dose for therapeutic procedures in infants <5 kg (4.07 vs 0.80 Gy·cm2). While weight predicts diagnostic dose, complexity drives therapeutic exposure. These benchmarks enable local optimization and support future National DRL development.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.